Owner's Operator and
Maintenance Manual
DEMAND OXYGEN DELIVERY
DEVICE AND SYSTEM
Venture 20EX, 20 PSIG Device
Venture 50EX, 50 PSIG Device
Venture Pack 211EX, 20 PSIG Shoulder Bag System
w/M6 Cylinder
Venture Pack 212EX, 20 PSIG Shoulder Bag System
w/M9 Cylinder
DEALER: THIS MANUAL MUST BE GIVEN TO THE USER OF THIS
DEMAND OXYGEN DELIVERY DEVICE.
USER: BEFORE USING THIS DEMAND OXYGEN DELIVERY
DEVICE, READ THIS MANUAL AND SAVE FOR FUTURE REFERENCE.
TABLE OF CONTENTS
SAFETY SUMMARY ............................................. 4
UNPACKING/HANDLING ..................................... 8
INTRODUCTION ................................................ 10
Invacare Venture Demand Oxygen
Delivery Device ....................................................................................... 10
FEATURES ......................................................... 11
Venture Demand Oxygen Delivery Device
Components and Definition of Terms ................................................. 11
Venture Demand Oxygen Delivery Device System
Components and Definition of Terms ................................................. 13
SPECIFICATIONS ............................................... 14
OPERATION ...................................................... 16
The Use of Oxygen.................................................................................... 16
Set-up and Normal Pulse Operation .................................................... 16
MAINTENANCE................................................. 20
Prolonging the Life of Your Battery....................................................... 20
Care of Your Venture Demand Oxygen Delivery
Device and System.................................................................................... 21
Cleaning the Outer Case.......................................................................... 21
Carrying Case .............................................................................................. 21
ACCESSORIES AND SPARE PARTS ..................... 22
TROUBLESHOOTING.......................................... 23
OXYGEN CONSERVATION................................. 26
Battery Use Time........................................................................................ 26
Compatibility ............................................................................................... 26
LIMITED WARRANTY ........................................ 27
PART NO. 1093845
3
VENTURE
SAFETY SUMMARY
WARNING
This device should ONLY be used under the supervision
of a physician. Verify your oxygen flowrate with your
physician when you begin using your Demand Oxygen
Delivery System.
If you have any questions about the oxygen delivery
system or its use, call your Home Health Care Dealer or
Service Representative.
If you suspect that an oxygen delivery system has been
damaged, contact your Service Representative
IMMEDIATELY. Your Service Representative MUST
handle all maintenance.
Keep all oxygen delivery systems out of the reach of
children.
Avoid creation of any spark near medical oxygen
equipment. This includes sparks from static electricity
created by any type of friction.
All electrical equipment near oxygen units MUST be
properly grounded.
This system only delivers medical grade (U.S.P.)
oxygen when used with oxygen cylinders marked as
such. When using this system with Invacare Venture
Homefill, oxygen delivery will be USP 93%+3%. Oxygen
supplied by the system is supplemental and is NOT to
be considered life-supporting.
Before using the Demand Oxygen Delivery Device, check
the contents gauge of the oxygen supply
equipment. The oxygen supply equipment MUST be filled
only by an authorized Service Representative, unless
being filled by the Invacare Venture Homefill system.
If at anytime the attendant or the end user concludes
that an insufficient amount of oxygen is being received,
VENTURE
4
PART NO. 1093845
SAFETY SUMMARY
WARNING (CONTINUED)
the supplier and/or physician should be contacted
IMMEDIATELY. No adjustments should be made to
the flowrate unless prescribed by a physician.
DO NOT attempt to alter the switches in the battery
compartment. These switches are preset by your
Home Health Care Dealer and control the breath
delivery prescription.
RISK OF FIRE:
Oxygen does not burn. However, it will accelerate
burning of flammable materials.
TO PREVENT FIRE:
DO NOT SMOKE when operating the oxygen delivery
system. No Smoking signs should be prominently
displayed.
Keep oxygen delivery system at least 5-feet away from
ANY electrical appliances; such as radios, TVs, air
conditioners, fans, electric razors, etc.
Keep the oxygen delivery system away from open
flames or heating sources such as furnaces or stoves.
Keep flammable materials away from the oxygen
delivery system; such as oils, grease and aerosol sprays.
These will ignite and burn quickly in the presence of
concentrated oxygen.
TO PREVENT HIGH CONCENTRATIONS OF OXYGEN:
Keep the oxygen delivery system in a well ventilated
area at all times.
When the oxygen delivery system is not in use, press
the OFF button to place the Demand Oxygen Delivery
Device in the Continuous Mode, turn the Cylinder OFF,
bleed the remaining oxygen from the Regulator, turn
the Regulator OFF.
PART NO. 1093845
5
VENTURE
SAFETY SUMMARY
WARNING (CONTINUED)
DO NOT wear the Demand Oxygen Delivery Device
under any clothing.
OTHER PRECAUTIONS:
Never attempt to lubricate the oxygen delivery system.
DO NOT insert any object (except suitable AC/DC
Adapter) in external power jack of Demand Oxygen
Delivery Device. This may cause an electrical shock
which could cause burns.
DO NOT subject the Demand Oxygen Delivery Device to
temperatures below -40°F (-40°C) or above 158°F (70°C).
To do so could damage the electronics and battery.
NEVER tamper with the oxygen delivery system.
Doing so could create a hazardous condition and
possibly cause the oxygen delivery system to malfunction.
If you have any questions, call your Home Health Care
Dealer or Service Representative.
ALWAYS connect the Demand Oxygen Delivery
Device to a pressure regulated oxygen source.
ONLY connect 20 psig Demand Oxygen Delivery
Device to a 20 psig oxygen source and a 50 psig
Demand Oxygen Delivery Device to a 50 psig oxygen
source.
DO NOT use any additional lengths of tubing between the
oxygen source and the Demand Oxygen Delivery Device.
ONLY use Invacare tubing supplied with the Demand
Oxygen Delivery Device, if any tubing is required.
DO NOT use a mask when using pulse mode delivery,
USE ONLY a standard nasal cannula.
When using the continuous flow delivery, a mask or
nasal cannula can be used.
VENTURE
6
PART NO. 1093845
SAFETY SUMMARY
WARNING (CONTINUED)
DO NOT use a humidifier with the Demand Oxygen
Delivery Device.
DO NOT expose the Demand Oxygen Delivery Device
or system to water or any other liquids.
RADIO FREQUENCY INTERFERENCE
Most electronic equipment is influenced by Radio Frequency
Interference (RFI). CAUTION should be exercised with
regard to the use of portable communications equipment in
theareaaroundsuchequipment.
PART NO. 1093845
7
VENTURE
UNPACKING / HANDLING
UNPACKING
1. Check for any obvious damage to the carton or its contents.
(If damage is evident, notify your local Invacare Home Health
Care Dealer/Carrier.)
2. Remove all the loose packing from the carton.
3. Carefully remove all the components from the carton.
The following is a listing of items included in the device and
system packages:
VENTURE50EX,
MODELIDD50EX
50 psig Demand Device with
Threaded Fitting Battery
Venture Operator's Manual.
VENTURE20EX,
MODELIDD20EX
20 psig Demand Device with Barb
Fitting and Connection Tubing
Battery
Venture Operator's Manual Optional
Threaded Fitting.
VENTURE PACK 211EX,
MODELIDD211EX
IDD20EX Demand Device With
Connecting Tube
AND VENTURE PACK 212EX,
MODELIDD212EX
20 psig Piston Type Flow Control
Regulator (0-6 L/min.)
Carrying Case: Dual Size Hunter
Green Shoulder Bag
Cylinder:
M6 Cylinder (IDD211EX)
M9 Cylinder (IDD212EX)
Nasal Cannula, Cylinder Wrench,
VENTURE
8
PART NO. 1093845
HANDLING PROCEDURES
NOTE: Unless the Venture Demand Oxygen Delivery Device is to be
used immediately, retain boxes, containers and packing materials
for use in returning or storing the equipment until usage is required.
INSPECTION
1. Examine exterior of the equipment for nicks, dents,
scratches or other damages. Inspect all components.
STORAGE
1. Carefully repackage all components in the original packaging.
2. Store the repackaged equipment in a dry area.
3. DO NOT place other objects on top of the repackaged
equipment.
PART NO. 1093845
9
VENTURE
INTRODUCTION
INVACARE VENTURE DEMAND OXYGEN
DELIVERY DEVICE
Ordinary oxygen supply units deliver oxygen continuously.
The normal breathing pattern is to inhale about one-third of the time,
and exhale two-thirds of the time. This means that continuous flow
units may be inefficient during user exhalation.
In contrast, the Venture Demand Oxygen Delivery Device (Venture
DODD) senses your inhalation and instantly releases a programmed
pulse of oxygen at a specified flowrate at the beginning of inhalation.
The specified flowrate of oxygen is delivered within the first second of
inhalation. Oxygen is not delivered during exhalation. This means that
your oxygen use time is extended by the ratio of breath duration over
the pulse duration. Venture DODD will extend user time an average of
four (4) times in the Variable duration breath mode.
Comfort is another advantage of the Venture DODD. Nasal drying caused
by continuous flow is eliminated, as oxygen is only delivered during
inhalation. Therefore, additional humidification is not necessary. Since the
oxygen is delivered at normal flowrates, there is not a high blast of oxygen
to the end user, which makes the Venture DODD more comfortable than
other pulse systems.
The Venture DODD can be used in the traditional continuous flow mode
simply by turning the Venture DODD device OFF. Effective flow deliveries of
0.5 to 6 liters per minute (L/min.) can be instantly selected, in pulse or
continuous mode, by each individual as prescribed by your physician.
During pulse mode, the Venture DODD runs on a disposable D-cell battery
which must be replaced after approximately 80 hours of normal use
(Refer to the OPERATION section in this manual for information on the
Low Battery Indicator LED). During continuous mode operation, the unit
does not need the battery to operate.
NOTE: Actual Battery use time depends on your breathing rate.
The Venture DODD can be used with compressed gas or liquid
oxygen systems. Both 20 and 50 psig units were designed for use on
20 and 50 psig oxygen delivery systems, respectively.
NOTE: The 50 psig system should ONLY be used with the threaded (DISS)
fitting, and NOT the 12-inch connecting tube, as the pressure may be too
great for the tube to stay connected.
VENTURE
10
PART NO. 1093845
FEATURES
VENTURE DEMAND OXYGEN DELIVERY
DEVICE COMPONENTS AND DEFINITION
OF TERMS (FIGURE 1)
ON/OFF Buttons - Determines flow selection method. The ON
button turns Power ON for normal pulse operation (
the OFF button disengages the electronics for continuous flow
operation ( ).
) and
NOTE: The OFF button does NOT stop flow delivery and is NOT
recommended for unit shutdown. See SETUP AND NORMAL PULSE
OPERATION in the OPERATION Section of this manual.
Oxygen OUTLET Fitting - The delivery point of oxygen to the
end user from the Venture DODD. The fitting is barbed to
securely hold the cannula for oxygen delivery.
Oxygen INLET Fitting - The delivery point of oxygen from the
oxygen system to the Venture DODD.
External Power Input Jack - AC/DC Power Adapter connection port.
Status Indicators:
O2 - GREEN LED will illuminate with each breath indicating
that the Venture DODD is operating.
- Attention - YELLOW LED illumination indicates the
Venture DODD is NOT sensing end user's inhalation or
that there is an internal malfunction.
Refer to TROUBLESHOOTING Section.
- Low Battery - RED LED indicates two (2) low battery
levels.
O2 - Low/No Gas - GREEN LED illumination indicates
Venture DODD is NOT sensing O2 presence in pulse
delivery.
PART NO. 1093845
11
VENTURE
FEATURES
External Power
Input Jack
ç
é
é
Oxygen OUTLET Fitting Oxygen INLET Fitting
Low O2 LED (GREEN)
Battery LED (RED)
Attention LED (YELLOW)
Pulse LED (GREEN)
êêêê
ON(Pulse ModeOperation)
è
é
OFF (Continuous Mode Operation)
NOTE: Battery Compartment located on
bottom back of Venture Demand Oxygen
Delivery Device.
è
FIGURE 1 - VENTURE DODD FEATURES
VENTURE
12
PART NO. 1093845
FEATURES
VENTURE DODD SYSTEM COMPONENTS AND
DEFINITION OF TERMS (FIGURE 2)
Cylinder - Lightweight aluminum cylinder provided.
Regulator - Provides flowrate control of oxygen delivery with
compressed gas cylinders.
DISS Fitting - Connection point for direct connect to oxygen source.
Barb Fitting - Connection point for tubing used with compressed
gas cylinders.
Connection Tubing - Provides oxygen path from regulator to
the device inlet barb fitting for use with compressed gas cylinders.
Carrying Case with Padded Strap - Used for convenient
transportation and is adjustable to ensure comfort.
Optional AC/DC Adpater - Designed to provide AC current to device.
Barb Fitting
Regulator
DISS Fitting
M9 Cylinder
Tubing
M6 Cylinder
Case
Optional AC/DC
Adapter
(Cannula and Wrench not shown)
FIGURE 2 - VENTURE DODD SYSTEM FEATURES
PART NO. 1093845
13
VENTURE
SPECIFICATIONS
OPERATING VOLTAGES
EXTERNAL AC/DC POWER
ADAPTER:
Output: 12 + 3 VDC, 100 mA
Connector: 5.5mm OD x 2.1mm
ID x 12.0mm No. 2 Connector
ALARMS
Low Battery (Two [2] levels)
Cannula Disconnect (Breath not
sensed)
Device Malfunction
Low / No Gas Indicator
INDICATORS/SYMBOLS:
O2 LED:
Pulse Delivery
LED:
LED:
Cannula Disconnect/Missing
Breath Indicator, Audible Alarm
Battery Low Charge Indicator (1/2 to 2
hours of use time left), Audible Alarm
O2
LED:
Low or No Gas, Audible Alarm
Class BF
:
BATTERY:
Approx. Battery Usage Time:
Alkaline: (1) "D" size cell
80 Hours @ 20 bpm
(Variable Breath Mode)
OXYGEN DOSE DURATION:
1-variable duration breathing:
Auto-adjusts between 0.5 and 1.0
seconds based on patient breathing rate.
3-fixed duration breath mode:
Fixed 1.0 seconds, 0.75 seconds,
0.50 seconds
FLOWRATERANGE(L/Min.):
0 to 6 L/Min.
OXYGENSOURCEPRESSURE:
Venture 20:
Venture 50:
20 + 3 psig (138 + 20 kPa)
50 + 5 psig (345 + 35 kPa)
REGULATOR
Flow Settings:
0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0,
5.0, 6.0 L/Min.
Regulated Pressure:
20 psig
VENTURE
14
PART NO. 1093845
SPECIFICATIONS
ENVIRONMENTAL
Operating Temperature:
Storage Temperature:
40° to 104° F (5° to 40°C)
- 40° to 158°F (-40° to 70°C)
DEVICE DIMENSIONS
Length:
Thickness:
Width:
Weight:
5.4-inches (13.7cm)
1.6-inches (4.0cm)
3.6-inches (9.1cm)
16.0 ozs. (454gm)
PART NO. 1093845
15
VENTURE
OPERATION
THE USE OF OXYGEN
The type of oxygen administrating equipment to be used, as well as the
flowrate and time of usage, must be prescribed by a physician. Your
physician should verify your Venture DODD flowrate prescription.
WARNING
If at anytime the attendant or the end user concludes
that an insufficient amount of oxygen is being received,
the supplier and/or physician should be contacted im-
mediately. No adjustments should be made to the
flowrate unless prescribed by a physician.
NOTE: ONLY a NASAL CANNULA can be used in the PULSE MODE to
deliver oxygen. An oxygen mask or nasal cannula can be used with
Continuous Flow to deliver oxygen. Low flow cannula should not be used.
SET-UP AND NORMAL PULSE OPERATION
1. Install the battery into the Venture DODD or plug AC/DC
power adapter into external power jack.
A. To Install or Replace the Battery:
Open the battery cover on the back of the Venture
DODD by applying light inward pressure with your thumb
to the cover latch and then gently lift the cover open.
If replacing a battery: Remove the used battery from the
battery compartment by grasping the top and bottom ends
and pulling straight out.
To install the battery: Orient the new battery so that its
polarity matches the polarity marked in the battery compartment.
WARNING
DONOTattempttoaltertheswitchesinthebattery
compartment.Theseswitchesarepresetbyyourhome
healthcareproviderandcontrolthebreathdeliveryprescription.
Press the new battery in place and close the battery cover.
2. Check Oxygen Source to verify adequate oxygen supply.
VENTURE
16
PART NO. 1093845
OPERATION
3. If using an oxygen cylinder, check that the regulator is
properly attached before opening the cylinder valve.
4. Connect the Venture DODD to the regulator by performing
one (1) of the following:
A. If using an Oxygen Cylinder and Regulator:
On a 20 psig regulator, connect the tubing (supplied with
the Venture DODD) from the barb fitting of the regulator
to the barb inlet fitting on the Venture DODD.
On a 50 psig regulator, connect the threaded (DISS) fitting
(supplied with the Venture DODD) to the regulator DISS fitting.
NOTE: The 50 psig system should ONLY be used with the threaded
(DISS) fitting and not the 12-inch connecting tube, as the pressure may
be too great for the tube to stay connected.
B. If using a Liquid Oxygen Reservoir:
Directly connect the Venture DODD to the oxygen
source using a DISS fitting.
5. Connect the nasal cannula tubing to the barbed oxygen outlet
fitting on the Venture DODD and adjust cannula for comfort.
WARNING
When using nasal or masked type administering
equipment, oxygen or air tubing MUST be routed and
secured properly to ensure that the tubing does NOT
become entangled, kinked or severed.
NOTE: Use ONLY a NASAL CANNULA with PULSE MODE
delivery, NOT a mask. A mask may NOT fit your face well enough for the
unit to sense your inhalation.
NOTE: A special nasal cannula is NOT required. Any standard flow
crush-proof nasal cannula can be used.
NOTE: A MASK or a CANNULA may be used with CONTINUOUS FLOW.
NOTE: The cannula tubing can be up to 50 feet long to function properly
in the pulse mode. The continuous flow mode is NOT affected by the
length of the canula tubing.
PART NO. 1093845
17
VENTURE
OPERATION
6. Set the prescribed flow rate on the oxygen source by performing
one (1) of the following:
A. If using an Oxygen Cylinder and Regulator:
OPEN cylinder valve and adjust regulator to prescribed flow rate.
B. If using a Liquid Oxygen Reservoir:
Set flow control knob on reservoir to prescribed flowrate.
7. Press the ON button to engage pulse mode. The unit will
pulse each time the user inhales, the GREEN PULSE LED will
blink with each oxygen pulse delivery.
8. As battery power depletes, the unit will alarm to inform the end
user to replace the battery or switch to external power source.
A. INITIAL BATT. LED - The Red LED will FLASH and the
audible alarm will BEEP three (3) times. This indicates that
you have about two (2) hours of use time left on your battery.
B. SECOND BATT. LED - The Red LED will be SOLID and
the audible alarm will BEEP six (6) times. This indicates that you
have between 15 and 30 minutes of use time left on your battery.
NOTE: Remaining battery use times are dependent on battery type
being used. Alkaline battery type is recommended for longest life.
OFF
(Continuous Mode
Operation)
FIGURE 3 - SWITCHING TO CONTINUOUS FLOW
VENTURE
18
PART NO. 1093845
OPERATION
NOTE: Whenever the pulse mode is used, the battery is monitored.
If the SECOND BATT. LED illuminates, either replace the battery or
switch to continuous mode if oxygen supply is adequate. When in
continuous mode, the battery is NOT used.
NOTE: When in the continuous flow mode, the oxygen supply will be
used considerably faster then in the pulse mode. If using a portable
oxygen supply, make plans for an alternative source of oxygen. Replace
the battery as soon as possible.
9. If it is necessary to switch to continuous flow at any time,
press the OFF button (FIGURE 3).
10. If turning the system OFF for cylinder replacement or storage,
press the Venture DODD OFF button. Turn OFF the cylinder
post valve, bleed the oxygen from the regulator and turn OFF
the regulator prior to any disassembly.
PART NO. 1093845
19
VENTURE
MAINTENANCE
PROLONGING THE LIFE OF YOUR BATTERY
The Venture DODD battery is powered by a single D-cell battery.
The Alkaline battery type is recommended for longest life.
NOTE: The Alkaline battery should last for approximately 80 hours with
normal use, but will last even longer if you follow this procedure:
1. Use the battery to its fullest capacity prior to replacing. This is
accomplished by running the unit from the battery until it stops
working (approximately 15 to 30 minutes after the SECOND
BATT. indicator illuminates SOLID and the audible alarm
sounds six [6] times).
CAUTION
DO NOT store Venture DODD with the battery installed.
WARNING
DO NOT recharge, dispose of in fire, mix with different
battery type, connect improperly or disassemble
Alkaline batteries. Doing so may cause battery to
explode or leak resulting in injury.
2. Operate your unit from the AC/DC power adapter when at home.
3. Avoid exposure to extreme temperatures.
WARNING
DO NOT operate the unit in temperatures above
104°F (40°C) or below 40°F (4°C), or place near heat
sources such as radiators or fires.
Electrical shock hazard. DO NOT remove outer case
of this unit. All disassembly and maintenance of this
unit MUST BE done by an Invacare Dealer. If outer
case is removed by anyone EXCEPT an Invacare
Dealer the warranty will be voided.
VENTURE
20
PART NO. 1093845
MAINTENANCE
CARE OF YOUR VENTURE DEMAND
OXYGEN DELIVERY DEVICE AND SYSTEM
CLEANING THE OUTER CASE
WARNING
DO NOT submerge in water to clean.
DO NOT open outer case. If outer case is removed by
anyone EXCEPT an Invacare Dealer the warranty will
be voided.
CAUTION
Use no oil. Never apply waxes, polishes, or other similar
compounds to any surface of the Venture Demand
Oxygen Delivery Device, as these compounds may be
flammable.
Clean outer case with damp cloth. If necessary, a mild dish
detergent may be used. Dry outer case with lint free cloth.
CARRYING CASE
The carrying case can be hand washed in warm water. Air dry
ONLY. NO bleach. DO NOT dryclean.
PART NO. 1093845
21
VENTURE
ACCESSORIES AND SPARE PARTS
M6 Aluminum Cylinder, 164L Capacity
M9 Aluminum Cylinder, 248L Capacity
Piston Regulator, 0.5-6 L/Min., 20 psig, CGA 870
Shoulder Bag, Hunter Green
(Dual size for Cylinder diameters of 3 and 4-inches)
Venture Device Pouch
12-inch Connection Tubing
Barb Fitting, 1/4-28 UNF
7-Foot (2.1m) Nasal Cannula
AC/DC Adapter
Venture Operator's Manual
NOTE: Rechargeable batteries can be used to power Venture DODD,
however the batteries and charger are a user supply item not provided
by Invacare.
NOTE: If you have any questions regarding accessories, please call
Invacare or your Home Healthcare Dealer.
VENTURE
22
PART NO. 1093845
TROUBLESHOOTING
PROBLEM
POSSIBLE CAUSE
SOLUTIONS
Indicates Oxygen Pulse Flow
in normal pulse delivery.
O2 LED - (GREEN)
Blinking.
This is normal
operation.
Indicates Venture DODD
not sensing presence of
oxygen in pulse delivery.
O2 LED - (GREEN)
Turn unit OFF, check
cylinder pressure is above
refill limit (500 psig) and
post valve is open, check
regulator is set to correct
flowrate. If problem still
exists, call your Home
Health Care Dealer or
Service Representative
IMMEDIATELY and go to
continuous mode if
oxygen supply is adequate.
Indicates oxygen pulse
delivery, but Venture DODD
not sensing inhalation.
O2 and
(GREEN and YELLOW)
Blinking
LED -
Breathe normally. Check
cannula for kinking and
check tubing connection.
Indicates no pulse
delivery, Venture DODD
not sensing inhalation.
Breathe normally. Check
cannula for kinking and
check tubing connection.
LED - (YELLOW)
Blinking and
Intermittent Beep
Possible internal problem.
End user possibly not
receiving oxygen.
LED - (YELLOW)
Solid and Continuous
Beep
Turn unit off then on
again. If problem still
exists, call your Home
Health Care Dealer or
Service Representative
IMMEDIATELY and go to
continuous mode if
oxygen supply is adequate.
LED - (RED)
Battery low.
Replace battery as soon
as possible.
Blinking, Three (3) Beeps Approximately two (2) hours
at Beginning.
of battery use time left.
Approximately 30 minutes
of battery use time left.
LED - (RED) Solid,
Six (6) Beeps at Beginning.
Replace battery or turn
unit OFF to enable
continuous mode if
oxygen supply is adequate.
PART NO. 1093845
23
VENTURE
TROUBLESHOOTING
PROBLEM
POSSIBLE CAUSE
SOLUTIONS
Possible internal sensor
problem. End user
possibly not receiving
oxygen.
Turn unit off then on
again. If problem still
exists, call your Home
Health Care Dealer or
Service Representative
IMMEDIATELY and go to
continuous mode if
LED - (RED) Solid,
Continuous Beep.
oxygen supply is adequate.
No lights illuminate or
RED and YELLOW LEDs
on together.
Electronic failure.
Call your Home Health
Care Dealer or Service
Representative
IMMEDIATELY and go
to continuous mode if
oxygen supply is
adequate.
Check contents gauge on
oxygen source. Call your
Home Health Care Dealer
or Service Representative
IMMEDIATELY.
Oxygen supply exhausted.
Oxygen NOT being
delivered.
Cylinders - OPEN valve.
Liquid Oxygen Supply -
Make sure coupling is
FULLY engaged.
Oxygen supply NOT
turned ON.
Check all cannula
connections.
Cannula kinked or
crushed.
Turn regulator knob to
prescribed flowrate.
Regulator knob NOT
turned ON.
Call your Home Health
Care Dealer or Service
Representative
Internal Problem.
IMMEDIATELY.
External input will not
power device.
Push plug in all the way.
Connector jack not fully
inserted into unit.
Wall outlet bad or no
power.
Check that outlet power
switch is on. Check outlet
by plugging in another
appliance or lamp.
VENTURE
24
PART NO. 1093845
TROUBLESHOOTING
PROBLEM
POSSIBLE CAUSE
SOLUTIONS
External input will not
power device (Cont.).
Incorrect power adapter.
Check that power adapter
used within specification for
operating voltage of external
power supply. Call your
Home Health Care Dealer
or Service Representative
IMMEDIATELY.
Check battery
"Low" battery light
Battery life expired.
manufacturer's
comes on too soon after
battery replacement.
expiration date. See
Prolonging the Life of the
Battery in the MAINTE-
NANCE Section of this
manual. Call your Home
Health Care Dealer or
Service Representative
IMMEDIATELY.
External power adapter
plug does not go into
jack in unit.
Jack or plug damaged.
Call your Home Health
Care Dealer or Service
Representative
IMMEDIATELY.
Non-compatible adapter
plug.
Refer to SPECIFICATION
Section for external
power adapter connector
size.
Check if YELLOW and
GREEN O2 LEDS illumi-
nate. If so, see indicator
description in this section.
Call your Home Health
Care Dealer or Service
Representative
Unit pulses at times
other than inhalation.
Unit may be in backup
Pulse Mode.
IMMEDIATELY.
Verify compatibility of
model and pressure.
Call your Home Health
Care Dealer or Service
Representative
Operating pressure of
oxygen supply source too
high.
Continuous flow out of
cannula when in pulse
mode.
IMMEDIATELY.
NOTE: If any of the above problems occur and the solutions do not work,
please contact your Invacare Home Health Care Dealer for service.
PART NO. 1093845
25
VENTURE
OXYGEN CONSERVATION
OXYGEN CONSERVED AND BATTERY
USE TIME BASED ON BREATH RATE
BREATH RATE (BPM)
% OXYGENCONSERVED
Variable Breath
Fixed 1.0 sec. Breath Mode:
Fixed 0.75
Fixed 0.50 sec. Breath Mode:
10
20
30
40
Mode:
8277
70
63
75
83
Breath
91
66
Mode:
83
50
75
33
87
66
sec.
i.e. 77% Oxygen conservation over Continuous Flow achieved
with Venture DODD in Variable Breath Mode @ 20 BPM Rate.
Oxygen supply lasts 4.3-times longer. Venture DODDresponds to
individual breath rates. Therefore, use times stated are only
approximates. Actual savings depends on individual breathing
patterns.
COMPATIBILITY
Venture DODD requires using a flowrate controlled oxygen
source. Gaseous or liquid oxygen systems that meet the
following requirements are compatible to the Venture
DODD:
Venture 20 requires a 20 +3 psig (138 + 20 kPa) flow source.
Venture 50 requires a 50 +5 psig (345 + 35 kPa) flow source.
Flowrate Range: 0-6 L/Min.
VENTURE
26
PART NO. 1093845
LIMITED WARRANTY
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED
TO COMPLY WITH FEDERAL LAW APPLICABLE TO
PRODUCTS MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser/user of
our products.
This warranty gives you specific legal rights and you may also have
other legal rights which vary from state to state.
Invacare warrants the Demand Oxygen Delivery Device to be free
from defects in materials and workmanship for a period of one (1)
year from date of purchase. In addition, all internal electronic
components and switches carry a three (3) year warranty from
date of purchase. If within such warranty period any such product
shall be proven to Invacare satisfaction to be defective, product
shall be repaired or replaced, at Invacare option. This warranty
does not include normal wear and tear or shipping charges
incurred in replacement part installation or repair of such
product. Invacare sole obligation and your exclusive remedy
under this warranty shall be limited to such repair and/or
replacement.
For warranty service, please contact Invacare Service Department
at the Toll Free Number listed below during normal hours.
Upon receiving notice of an alleged defect in a product, Invacare
will issue a serialized return authorization. It shall be the
responsibility of the purchaser to return, at the purchasers cost,
to either Invacares plant or service center as specified by Invacare
in advance. Defective components part(s) must be returned for
warranty inspection using the serial number for identification
within 30 days of the return authorization date. DO NOT return
products to our factory without prior consent. C.O.D. shipments
will be refused - please prepay shipping charges.
LIMITATIONSANDEXCLUSIONS:THEFOREGOING
WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PROD-
UCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DE-
FACED, PRODUCTSSUBJECTEDTONEGLIGENCE,
ACCIDENT, IMPROPEROPERATION, MAINTENANCEORSTOR-
AGE, PRODUCTSMODIFIEDWITHOUTINVACARESEXPRESS
WRITTENCONSENT(INCLUDING, BUTNOT
LIMITEDTO, MODIFICATIONTHROUGHTHEUSEOF
UNAUTHORIZEDPARTSORATTACHMENTS;PRODUCTS
DAMAGED BY REASON OF REPAIRS MADE TO ANY
COMPONENTWITHOUTTHESPECIFICCONSENTOF
INVACARE, OR TO A PRODUCT DAMAGED BY
CIRCUMSTANCESBEYONDINVACARERESPIRATORY -
PART NO. 1093845
27
VENTURE
LIMITED WARRANTY
CORPORATIONSCONTROL, ANDSUCHEVALUATIONWILLBE
SOLELYDETERMINEDBYINVACARE. THEWARRANTYSHALL
NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR
FAILURETOADHERETOTHEFOLLOWINGINSTRUCTIONS.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN
LIEU OF ANY OTHER WARRANTIES WHATSOEVER,
WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR
VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE
LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE
PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN.
THE APPLICATION OF ANY IMPLIED WARRANTY
WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION
OF THE EXPRESS WARRANTY PROVIDED HEREIN.
INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL
OR INCIDENTAL DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW THE EXCLUSION OR
LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE,
OR LIMITATION ON HOW LONG AN IMPLIED WARRANTY
LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY
NOT APPLY TO YOU.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH
STATE/PROVINCIAL LAWS AND REQUIREMENTS.
Invacare Corporation
_____________________________________________________________________________________________________________________
USA
One Invacare Way
Elyria, Ohio USA
44036-2125
Invacare is a registered trademark
of Invacare Corporation.
Yes, you can. is a trademark of
Invacare Corporation.
800-333-6900
©2002 Invacare Corporation
Canada
Part No. 1093845
Rev C 10/02
570 Matheson Blvd E Unit 8
Mississauga Ontario
L4Z 4G4 Canada
800-668-5324
|